Model Number M00565050 |
Device Problem
Premature Activation (1484)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 09/03/2015 |
Event Type
Injury
|
Manufacturer Narrative
|
Although the exact patient age was not provided, the patient was reported to be over 18 years of age.(b)(4).The complainant indicated that the device has been disposed and will not be returned for evaluation; therefore a failure analysis of the complaint device could not be completed.If any further relevant information is identified, a supplemental medwatch will be filed.
|
|
Event Description
|
It was reported to boston scientific corporation that a wallflex colonic stent was to be implanted to treat a malignant stricture in the colon during a colonoscopy with colonic stent placement procedure performed on (b)(6) 2015.Reportedly, the colon was not fully obstructed but the tumor would fully obstruct the colon within a couple weeks without a stent.According to the complainant, during the procedure while the nurse was deploying the wallflex colonic stent it was noted that the sheath was not retracting.The wallflex colonic stent misfired and deployed inside the patient significantly past the stricture.The wallflex colonic stent was removed from the patient with rat tooth forceps.The procedure was completed with a different device.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be fine.
|
|