• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC - GALWAY WALLFLEX¿ COLONIC; STENT, COLONIC, METALIC, EXPANDABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC - GALWAY WALLFLEX¿ COLONIC; STENT, COLONIC, METALIC, EXPANDABLE Back to Search Results
Model Number M00565050
Device Problem Premature Activation (1484)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 09/03/2015
Event Type  Injury  
Manufacturer Narrative
Although the exact patient age was not provided, the patient was reported to be over 18 years of age.(b)(4).The complainant indicated that the device has been disposed and will not be returned for evaluation; therefore a failure analysis of the complaint device could not be completed.If any further relevant information is identified, a supplemental medwatch will be filed.
 
Event Description
It was reported to boston scientific corporation that a wallflex colonic stent was to be implanted to treat a malignant stricture in the colon during a colonoscopy with colonic stent placement procedure performed on (b)(6) 2015.Reportedly, the colon was not fully obstructed but the tumor would fully obstruct the colon within a couple weeks without a stent.According to the complainant, during the procedure while the nurse was deploying the wallflex colonic stent it was noted that the sheath was not retracting.The wallflex colonic stent misfired and deployed inside the patient significantly past the stricture.The wallflex colonic stent was removed from the patient with rat tooth forceps.The procedure was completed with a different device.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be fine.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WALLFLEX¿ COLONIC
Type of Device
STENT, COLONIC, METALIC, EXPANDABLE
Manufacturer (Section D)
BOSTON SCIENTIFIC - GALWAY
Manufacturer (Section G)
BOSTON SCIENTIFIC - GALWAY
Manufacturer Contact
nancy cutino
100 boston scientific way
marlborough, MA 01752
5086834000
MDR Report Key5110718
MDR Text Key27022658
Report Number3005099803-2015-02760
Device Sequence Number1
Product Code MQR
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K061877
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Report Date 09/03/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Nurse
Device Expiration Date11/16/2016
Device Model NumberM00565050
Device Catalogue Number6505
Device Lot Number0017461015
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received Not provided
Initial Date FDA Received09/29/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/20/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-