• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOEHRINGER LABORATORIES, LCC AVELLE WND-FLO ACUTE UNIT; NEGATIVE PRESSURE WOUND THERAPY POWERED SUCTION PUMP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOEHRINGER LABORATORIES, LCC AVELLE WND-FLO ACUTE UNIT; NEGATIVE PRESSURE WOUND THERAPY POWERED SUCTION PUMP Back to Search Results
Model Number 446610
Device Problem Electro-Static Discharge (2149)
Patient Problem Electric Shock (2554)
Event Type  Injury  
Manufacturer Narrative
Convatec is submitting this report as a result of activities related to convatec remediation protocol (b)(4).Convatec is submitting this report pursuant to the provisions of 21cfr part 803.Any additional information received regarding this event after filing this report will be filed on a supplemental report (medwatch 3500a).
 
Event Description
It was reported that "nurse describes wearing a nylon gown frequently worn by staff.She was experiencing static electricity from the gown and upon turning off the pump, experienced a shock like sensation and saw a spark at the power knob and then pump stopped working.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AVELLE WND-FLO ACUTE UNIT
Type of Device
NEGATIVE PRESSURE WOUND THERAPY POWERED SUCTION PUMP
Manufacturer (Section D)
BOEHRINGER LABORATORIES, LCC
500 e. washington st
norristown PA 19403
Manufacturer Contact
jeanette johnson
211 american ave
greensboro, NC 27409
3362973009
MDR Report Key5111865
MDR Text Key27053247
Report Number2518417-2015-30050
Device Sequence Number1
Product Code OMP
Combination Product (y/n)N
PMA/PMN Number
K061788
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Nurse
Report Date 02/27/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Nurse
Device Model Number446610
Device Lot Number0904122
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/01/2011
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received Not provided
Initial Date FDA Received09/29/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/01/2009
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-