• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SORIN GROUP DEUTSCHLAND S5 ROLLER PUMP; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SORIN GROUP DEUTSCHLAND S5 ROLLER PUMP; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 10-80-00
Device Problem Device Issue (2379)
Patient Problem No Patient Involvement (2645)
Event Date 07/07/2015
Event Type  malfunction  
Manufacturer Narrative
Sorin group deutschland manufactures the s5 roller pump.The incident occurred in (b)(6).(b)(4).Sorin group received a report that the touch screen of the s5 roller pump experienced an anomaly.The issue was discovered by a sorin representative during preventative maintenance, so there was no patient involvement.The investigation is ongoing.A follow-up report will be sent when the investigation is complete.
 
Event Description
Sorin group received a report that the touch screen of the s5 roller pump experienced an anomaly.The issue was discovered by a sorin representative during preventative maintenance, so there was no patient involvement.
 
Manufacturer Narrative
There have been two field actions for this issue.The z-numbers are z-0956/0965-2011 and z-1149/1158-2012.Livanova (b)(4) manufactures the s5 roller pump.The event occurred in (b)(6).This medwatch report is being filed on behalf of livanova (b)(4).The livanova field service representative on site performing routine maintenance replaced the pump with a loaner.A second visit was performed where the service technician replaced the touch screen and the accessories with a new lcd touch screen.The unit was functionally tested without issue and returned to service.Considering the manufacturing year of the pump, livanova (b)(4) believes the most likely root cause is aging and wear of the part and "glude." this is a known issue for which a root cause investigation (b)(4) has been initiated.A review of the dhr did not identify any deviations or non-conformities relevant to the reported issue.Evaluated on site by livanova technician.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
S5 ROLLER PUMP
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
SORIN GROUP DEUTSCHLAND
lindberghstr. 25
munich, 80939
GM  80939
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich, 80939
GM   80939
Manufacturer Contact
joan ceasar
14401 w. 65th way
arvada, CO 80004
2812287260
MDR Report Key5116102
MDR Text Key27195708
Report Number9611109-2015-00405
Device Sequence Number1
Product Code DWB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K071318
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Other
Remedial Action Other
Type of Report Followup
Report Date 04/10/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number10-80-00
Device Catalogue NumberN/A
Device Lot NumberN/A
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received Not provided
Initial Date FDA Received10/01/2015
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received04/10/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/25/2007
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberZ-0956/0965-2011
Patient Sequence Number1
-
-