Model Number 10-80-00 |
Device Problem
Device Issue (2379)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 07/07/2015 |
Event Type
malfunction
|
Manufacturer Narrative
|
Sorin group deutschland manufactures the s5 roller pump.The incident occurred in (b)(6).(b)(4).Sorin group received a report that the touch screen of the s5 roller pump experienced an anomaly.The issue was discovered by a sorin representative during preventative maintenance, so there was no patient involvement.The investigation is ongoing.A follow-up report will be sent when the investigation is complete.
|
|
Event Description
|
Sorin group received a report that the touch screen of the s5 roller pump experienced an anomaly.The issue was discovered by a sorin representative during preventative maintenance, so there was no patient involvement.
|
|
Manufacturer Narrative
|
There have been two field actions for this issue.The z-numbers are z-0956/0965-2011 and z-1149/1158-2012.Livanova (b)(4) manufactures the s5 roller pump.The event occurred in (b)(6).This medwatch report is being filed on behalf of livanova (b)(4).The livanova field service representative on site performing routine maintenance replaced the pump with a loaner.A second visit was performed where the service technician replaced the touch screen and the accessories with a new lcd touch screen.The unit was functionally tested without issue and returned to service.Considering the manufacturing year of the pump, livanova (b)(4) believes the most likely root cause is aging and wear of the part and "glude." this is a known issue for which a root cause investigation (b)(4) has been initiated.A review of the dhr did not identify any deviations or non-conformities relevant to the reported issue.Evaluated on site by livanova technician.
|
|
Search Alerts/Recalls
|