Additional suspect medical device components involved in the event: model #: sc-3400-30, serial #: (b)(4), description: infinion splitter 2x8 kit(30cm).Sc-2316-50, (sn (b)(4)), all sixteen cables were fractured at the bent/kinked sections of the lead body, at apparent a click anchor site about 8 inches from the distal tip.The lead body was sharply bent at the end of the proximal.No cables were exposed.Sc-3400-30 (sn (b)(4)) device evaluation indicated that the splitter passed all performed test.Visual/x-ray inspections and impedance measurements were performed to ensure the device integrity.No anomalies were found.Sc-3400-30 (sn (b)(4)) as no anomalies or deviations were found during the complaint investigation site (cis) review, there is no reason to suspect a manufacturing defect as the source of the reported complaint.
|