Model Number HT70 PLUS |
Device Problem
Sticking (1597)
|
Patient Problem
No Patient Involvement (2645)
|
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).The covidien customer service engineer (cse) evaluated the device, and verified the customer reported malfunction.The ventilator is due for its two year maintenance (pm).However, since this pm is temporarily on hold (battery), this unit will be placed on the awaiting parts status, and will hold on replacing the single board computer (sbc) until all parts are available to be installed.The repair is not yet completed.
|
|
Event Description
|
It was reported that during the power on self-test (post), the ventilator got stuck at the six images screen.There was no patient involvement.
|
|
Manufacturer Narrative
|
(b)(4).The repair of the unit was completed, and the single board computer (sbc) printed circuit board (pcb) was replaced.The unit passed all testing and operates within the manufacturing specifications, and was returned to the customer.
|
|
Manufacturer Narrative
|
The device was repaired and the reported issue was isolated to interface between the device and the failed component.If information is provided in the future, a supplemental report will be issued.
|
|