• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NEWPORT MEDICAL INSTRUMENTS, INC. HT70 PLUS VENTILATOR; CONTINUOUS VENTILATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NEWPORT MEDICAL INSTRUMENTS, INC. HT70 PLUS VENTILATOR; CONTINUOUS VENTILATOR Back to Search Results
Model Number HT70 PLUS
Device Problem Sticking (1597)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The covidien customer service engineer (cse) evaluated the device, and verified the customer reported malfunction.The ventilator is due for its two year maintenance (pm).However, since this pm is temporarily on hold (battery), this unit will be placed on the awaiting parts status, and will hold on replacing the single board computer (sbc) until all parts are available to be installed.The repair is not yet completed.
 
Event Description
It was reported that during the power on self-test (post), the ventilator got stuck at the six images screen.There was no patient involvement.
 
Manufacturer Narrative
(b)(4).The repair of the unit was completed, and the single board computer (sbc) printed circuit board (pcb) was replaced.The unit passed all testing and operates within the manufacturing specifications, and was returned to the customer.
 
Manufacturer Narrative
The device was repaired and the reported issue was isolated to interface between the device and the failed component.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HT70 PLUS VENTILATOR
Type of Device
CONTINUOUS VENTILATOR
Manufacturer (Section D)
NEWPORT MEDICAL INSTRUMENTS, INC.
1620 sunflower avenue
costa mesa CA 92626
Manufacturer (Section G)
NEWPORT MEDICAL INSTRUMENTS, INC.
1620 sunflower avenue
costa mesa CA 92626
Manufacturer Contact
thom mcnamara
15 hampshire street
mansfield, MA 02048
5084524811
MDR Report Key5117914
MDR Text Key27570042
Report Number2023050-2015-00247
Device Sequence Number1
Product Code CBK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K111146
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Followup,Followup
Report Date 12/21/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberHT70 PLUS
Device Catalogue NumberHT70 PLUS
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/12/2015
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received Not provided
Initial Date FDA Received10/01/2015
Supplement Dates Manufacturer ReceivedNot provided
09/30/2015
Supplement Dates FDA Received12/15/2015
12/21/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/01/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-