Event Description
A handheld was returned due to no longer being needed.Analysis of the handheld was completed on 09/11/2015.During the analysis it was identified that the handheld was unable to advance past the welcome screen.The cause for the display anomaly is associated with resistance values being higher than expected in the touch screen circuitry.Once the display was replaced with a known good display, no further anomalies associated with the handheld performance were identified during the analysis.Analysis of the software was completed on 09/11/2015.No anomalies associated with flashcard software or databases were identified during the flashcard analysis.The flashcard and software performed according to functional specifications.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROGRAMMING SOFTWARE
Type of Device
PROGRAMMING COMPUTER
Manufacturer (Section D)
CYBERONICS - HOUSTON
100 cyberonics blvd
houston TX 77058
Manufacturer (Section G)
CYBERONICS - HOUSTON
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
njemile crawley
100 cyberonics blvd
suite 600
houston, TX 77058
2812287200
MDR Report Key5118342
MDR Text Key27587517
Report Number1644487-2015-05991
Device Sequence Number1
Product Code LYJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Report Date 09/11/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMODEL 250
Device Lot Number1073772
Other Device ID NumberVERSION 8.1
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/20/2015
Is the Reporter a Health Professional? No
Event Location Other
Initial Date Manufacturer Received Not provided
Initial Date FDA Received10/01/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/10/2010
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-