It is indicated that the product is not returning for evaluation.Therefore, a review in-house testing was performed.In-house strip testing on strip lot 332831 had met criteria.The product performed as expected.A review of the manufacturing records for the lot did not uncover any relevant non-conformances.Lot meets release specification.The customer was using expired strips.This may have contributed to the unexpected results experienced by the customer.Based on the information available, there is no indication of a product deficiency.No corrective action is required at this time.
|