Model Number 5392
Device Problems Nonstandard Device (1420); Connection Problem (2900)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 07/01/2015
Event Type  malfunction  
Manufacturer Narrative
The information submitted reflects all relevant data received.If additional relevant information is received, a supplemental report will be submitted.Product event summary: analysis confirmed the customer comment that the output connector assembly was broken and it was further noted that the battery drawer would not open without batteries in it when the upper and lower cases were together.All found defective parts were replaced and all other identified issues were resolved.(b)(4).
 
Event Description
It was reported that the generator was being sent in for its calibration and it was additionally reported that there was a chip out of the ventricular connector.It was requested that the calibration be covered under warranty.The generator was returned for service.No patient complications have been reported as a result of this event.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Type of Device
PULSE-GENERATOR, PACEMAKER, EXTERNAL
Manufacturer (Section D)
PLEXUS MANUFACTURING SDN. BHD
bayan lepas free industrial zo
bayan lepas 11900
MY  11900
Manufacturer (Section G)
MEDTRONIC CARDIAC RHYTHM HEART FAILURE
8200 coral sea st ne
mounds view MN 55112
Manufacturer Contact
anne schilling
8200 coral sea st ne
mounds view, MN 55112
7635052036
MDR Report Key5121414
MDR Text Key27646435
Report Number3004593495-2015-00283
Device Sequence Number1
Product Code DTE
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K150246
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other Health Care Professional
Remedial Action Recall
Type of Report Followup
Report Date 07/01/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number5392
Device Catalogue Number5392
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/07/2015
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received Not provided
Initial Date FDA Received10/02/2015
Supplement Dates Manufacturer Received07/01/2015
Supplement Dates FDA Received09/18/2017
Was Device Evaluated by Manufacturer? No
Date Device Manufactured12/02/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberZ-1356-2015
Patient Sequence Number1
-
-