Model Number D314TRG
Device Problems Premature Discharge of Battery (1057); Output above Specifications (1432)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 07/02/2015
Event Type  Injury  
Manufacturer Narrative
The information submitted reflects all relevant data received.If additional relevant information is received, a supplemental report will be submitted.Product event summary: the device was returned and analyzed.Analysis of the device revealed normal battery depletion.Concomitant product(s): 4193-78 lead, implanted: (b)(6) 2005; a 6725 lead adapter, implanted: (b)(6) 2013.(b)(4).
 
Event Description
It was reported that the device was replaced due to elective replacement indicator (eri).The battery longevity was less than expected.The device also had a pacing problem of high right ventricular (rv) threshold.The device was removed and a new device was implanted.No patient complications have been reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROTECTA XT CRT-D
Type of Device
DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO
Manufacturer (Section D)
MEDTRONIC MED REL MEDTRONIC PUERTO RICO
ceiba norte industrial park, r
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC CARDIAC RHYTHM HEART FAILURE
8200 coral sea st ne
mounds view MN 55112
Manufacturer Contact
anne schilling
8200 coral sea st ne
mounds view, MN 55112
7635052036
MDR Report Key5121536
MDR Text Key27427427
Report Number3004209178-2015-19740
Device Sequence Number1
Product Code NIK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010031
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Report Date 07/02/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date07/28/2014
Device Model NumberD314TRG
Device Catalogue NumberD314TRG
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/27/2015
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received Not provided
Initial Date FDA Received10/02/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/04/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age00073 YR
-
-