Model Number 5392 |
Device Problem
Device Displays Incorrect Message (2591)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 08/26/2015 |
Event Type
malfunction
|
Manufacturer Narrative
|
The information submitted reflects all relevant data received.If additional relevant information is received, a supplemental report will be submitted.(b)(4).
|
|
Event Description
|
It was reported that the external pulse generator experienced an error.The epg was returned for repair.There was no patient involvement.
|
|
Manufacturer Narrative
|
Product event summary: analysis confirmed the customer comment of an error and further noted that the battery wire insulation was pinched and the wire exposed, the display wires were pinched and the wire exposed and the output connector assembly was broken.It was also noted that the device failed incoming functional testing due to the printed circuit board (pcb) being misaligned, the main pcb was realigned and it passed its retesting.All found defective parts were replaced and all other identified issues were resolved including installing the updated battery contact design.
|
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|