Model Number 5392 |
Device Problem
Electrical /Electronic Property Problem (1198)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 07/06/2015 |
Event Type
malfunction
|
Manufacturer Narrative
|
This report is based solely on device analysis.No information to suggest a device related adverse event or product problem was received.If additional relevant information is received, a supplemental report will be submitted.Product event summary: the device failed incoming functional testing.The main printed circuit board (pcb) was realigned and the functional testing was rerun but it failed again.The main pcb was swapped out and when functional testing was again rerun the device passed.It was noted that the main pcb was out of specification in an electrical manner.It was further noted that the keypad had a soft lock/unlock button.All found defective parts were replaced and all other identified issues were resolved.(b)(4).
|
|
Event Description
|
The external pulse generator was returned in accordance with instructions affiliated with the current corrective field action and subsequently tested out of specification during manufacturer's analysis.There was no patient involvement.
|
|
Manufacturer Narrative
|
This device was reported as included in the field action noted but returned product investigation found the device did not perform as described in the field action.The device is no longer included as part of the field action.Further analysis was performed on the main board.Visual inspection revealed no anomalies.Bench analysis found that the board failed on a functional tester on the atrial channel for output noise.However additional analysis at the bench did not confirm the failure.Conclusion: the failure seen at the test console could not be confirmed.
|
|
Search Alerts/Recalls
|