SORIN GROUP DEUTSCHLAND S5 DOUBLE HEAD PUMP; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 10-85-00
Device Problem Device Issue (2379)
Patient Problem No Patient Involvement (2645)
Event Date 09/04/2015
Event Type  malfunction  
Manufacturer Narrative
Patient information was not provided.Sorin group (b)(4) manufactures the s5 double head pump.The incident occurred in (b)(6).This medwatch report is being filed on behalf of sorin group (b)(4).Sorin group received a report that the pump head for the s5 roller pump rotated irregularly.There was no report of patient injury.The investigation is ongoing.A follow-up report will be sent when the investigation is complete.
 
Event Description
Sorin group received a report that the pump head for the s5 double head pump rotated irregularly.There was no report of patient injury.
 
Manufacturer Narrative
(b)(4) manufactures the s5 double head pump.The incident occurred in (b)(6).This medwatch report is being filed on behalf of livanova (b)(4).A livanova field service representative was dispatched to the facility to investigate.The service representative was able to confirm the reported issue and traced the failure to a defective pump head.The pump head was replaced and the issue was resolved.The replaced pump head was returned to livanova (b)(4) for further investigation.During the investigation, it was found that the bearing was damaged and fluid was found inside the pump a review of the dhr did not identify any deviation or non-conformities relevant to the reported issue.Sorin group (b)(4) will continue to monitor for trends related to this type of issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
S5 DOUBLE HEAD PUMP
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
SORIN GROUP DEUTSCHLAND
lindberghstr. 25
munich, 80939
GM  80939
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich, 80939
GM   80939
Manufacturer Contact
joan ceasar
14401 w. 65th way
arvada, CO 80004
2812287260
MDR Report Key5126264
MDR Text Key27419097
Report Number9611109-2015-00389
Device Sequence Number1
Product Code DWB
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K071318
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Health Professional
Remedial Action Other
Type of Report Followup
Report Date 02/20/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number10-85-00
Device Catalogue NumberN/A
Device Lot NumberN/A
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received Not provided
Initial Date FDA Received10/02/2015
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received02/20/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/10/2009
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-