Model Number HT70P
Device Problems Sticking (1597); Failure of Device to Self-Test (2937)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The device was returned for investigation.The complaint was verified.The single board computer (sbc) board was replaced and the device then passed all testing.
 
Event Description
It was reported that during the power on self test (post), a ventilator display was stuck.There was no patient involvement.
 
Manufacturer Narrative
The device was repaired and the reported issue was isolated to interface between the device and the failed component.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HT70 PLUS VENTILATOR
Type of Device
CONTINUOUS VENTILATOR
Manufacturer (Section D)
NEWPORT MEDICAL INSTRUMENTS, INC
1620 sunflower avenue
costa mesa CA 92626
Manufacturer (Section G)
NEWPORT MEDICAL INSTRUMENTS, INC.
1620 sunflower avenue
costa mesa CA 92626
Manufacturer Contact
thom mcnamara
15 hampshire street
mansfield, MA 02048
5084524811
MDR Report Key5133622
MDR Text Key27966331
Report Number2023050-2015-00253
Device Sequence Number1
Product Code CBK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K111146
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Followup
Report Date 12/20/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberHT70P
Device Catalogue NumberHT70P
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer09/19/2015
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received Not provided
Initial Date FDA Received10/07/2015
Supplement Dates Manufacturer Received09/10/2015
Supplement Dates FDA Received12/20/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/01/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-