• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SIEMENS HEALTHCARE DIAGNOSTICS CLINITEK STATUS+; CT STATUS+

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SIEMENS HEALTHCARE DIAGNOSTICS CLINITEK STATUS+; CT STATUS+ Back to Search Results
Catalog Number 10379675
Device Problems Failure to Power Up (1476); Smoking (1585)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/15/2015
Event Type  malfunction  
Manufacturer Narrative
Siemens is in process of investigating customer returned connect platform.The cause for the event is unknown at this moment.
 
Event Description
Customer reported that when they went to power on the instrument for the first time, it would not power on.They put 6 aa duracell batteries in the instrument with the connect ((b)(4)) and the instrument started to smoke up from the bottom and up the sides.There was no report of injury due to this event.
 
Manufacturer Narrative
Investigation of the instrument determined that the root cause was a catastrophic failure of d16 on the main pcb.D16 is the overvoltage protector diode, which blows if the voltage applied to instrument is too high.Customer's instrument was replaced and customer is now operational.There is no evidence that a siemens product is not meeting specifications.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLINITEK STATUS+
Type of Device
CT STATUS+
Manufacturer (Section D)
SIEMENS HEALTHCARE DIAGNOSTICS
511 benedict avenue
tarrytown 10591
Manufacturer (Section G)
SIEMENS HEALTHCARE DIAGNOSTICS MANUFACTURING, LTD.
northern road
chilton industrial estate
sudbury CO10 2XQ
UK   CO10 2XQ
Manufacturer Contact
steven andberg
2 edgewater drive
norwood, MA 02062
7812693655
MDR Report Key5134786
MDR Text Key28172064
Report Number1217157-2015-00144
Device Sequence Number1
Product Code JIL
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K091216
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Other Health Care Professional
Type of Report Followup
Report Date 02/22/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Catalogue Number10379675
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received Not provided
Initial Date FDA Received10/08/2015
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received02/22/2016
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Type of Device Usage N
Patient Sequence Number1
-
-