Multiple attempts to obtain additional information and to replace the product were unsuccessful, but are ongoing.This issue with the damaged transformer housing for the pump in style device was addressed in investigation ir12-0037.The investigation found that the transformers are being damaged during shipment from the manufacturer to medela.This damage is causing the plastic housing to fail prematurely when subjected to normal use and foreseeable misuse.The packaging used by the manufacturer to ship the transformers to medela is not robust enough to handle all of the potential shipping, handling, and abuse conditions that could arise from logistics of the consolidation process.As a result of the investigation, the shipping and consolidation process has been modified to reduce the handling and potential for double stacking of the skids.The shipping packaging strength has also been increased to further protect the transformers during shipping.Complaints against this product are currently being monitored for effectiveness of the above mentioned corrective action.The product involved in the complaint was not returned for evaluation/investigation at this time.Therefore, no conclusions can be made as to the cause of the event.Should additional information or the original product be received, resulting in new, changed, or corrected information, a follow up report will be filed at that time.
|
The customer emailed on (b)(6) 2015, that her transformer to her pump in style advanced tote was a rev n, not a rev l as initially reported.The cause of the breach in the rev n transformer has not been determined at this time.It is currently being investigated under (b)(4).
|