• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SIEMENS HEALTHCARE DIAGNOSTICS INC STREAMLAB ANALYTICAL WORKCELL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SIEMENS HEALTHCARE DIAGNOSTICS INC STREAMLAB ANALYTICAL WORKCELL Back to Search Results
Model Number STREAMLAB ANALYTICAL WORKCELL
Device Problem Electrical Shorting (2926)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/15/2015
Event Type  malfunction  
Manufacturer Narrative
The customer contacted the siemens customer care center (ccc).Ccc instructed the customer to turn the decapper offline.A siemens customer service engineer (cse) was dispatched to the customer site.After evaluation of the instrument, the cse discovered the decapper micro controller had shorted out.The cse replaced the micro controller and reset the streamlab.The decapper is now properly functioning.The cause of smoke being emitted from the instrument was due to an electric short.The instrument is performing according to specifications.No further evaluation of the device is required.
 
Event Description
The operator of a streamlab core unit observed smoke being emitted from the decapper.The customer powered off the decapper.There were no reports of adverse health consequences due to the smoke emitted from the instrument.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STREAMLAB ANALYTICAL WORKCELL
Type of Device
STREAMLAB ANALYTICAL WORKCELL
Manufacturer (Section D)
SIEMENS HEALTHCARE DIAGNOSTICS INC
101 silvermine road
brookfield CT 06804
Manufacturer (Section G)
INPECO S.P. A
registration #: 3005509212
via givoletto 15
val della torre, 10040
IT  
Manufacturer Contact
margarita karan
511 benedict avenue
tarrytown, NY 10591
9145243105
MDR Report Key5140399
MDR Text Key27939735
Report Number1226181-2015-00576
Device Sequence Number1
Product Code LXG
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Health Professional
Report Date 09/15/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberSTREAMLAB ANALYTICAL WORKCELL
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received Not provided
Initial Date FDA Received10/09/2015
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-