DATEX-OHMEDA, INC. AESPIRE 7100; ANESTHESIA GAS MACHINE Back to Search Results
Device Problems Electrical /Electronic Property Problem (1198); Connection Problem (2900); Protective Measures Problem (3015)
Patient Problem No Patient Involvement (2645)
Event Date 09/22/2015
Event Type  malfunction  
Manufacturer Narrative
Ge healthcare¿s investigation into the reported occurrence is still ongoing.A follow-up report will be issued when the investigation has been completed.Block a: no report of patient involvement.The initial reporter is located outside the u.S., and therefore this information is not provided due to country privacy laws.Device evaluation anticipated, but not yet begun.
 
Event Description
The hospital reported that, during a preoperative checkout, the anesthesia machine was noted to shut down without an alarm.There was no report of patient involvement.
 
Manufacturer Narrative
A ge healthcare service representative performed a checkout of the equipment and confirmed the reported complaint.The service representative cleaned and reset all the connections on the control board, resolving the reported complaint.The unit was returned to service.The status of health professional is unknown and was accidentally selected in the initial report; the original reporter occupation is unknown and was accidentally selected in the initial report; the investigation revealed new information regarding the device codes.The above codes are to replace the device codes in the initial report.No further actions are planned at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AESPIRE 7100
Type of Device
ANESTHESIA GAS MACHINE
Manufacturer (Section D)
DATEX-OHMEDA, INC.
3030 ohmeda drive
madison WI 53718
Manufacturer (Section G)
DATEX-OHMEDA, INC.
3030 ohmeda drive
madison WI 53718
Manufacturer Contact
brad wheeler
3000 n. grandview blvd., w450
waukesha, WI 53188-1696
MDR Report Key5140861
MDR Text Key27958472
Report Number2112667-2015-00165
Device Sequence Number1
Product Code BSZ
Combination Product (y/n)N
PMA/PMN Number
K000706
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign
Type of Report Followup
Report Date 09/22/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received Not provided
Initial Date FDA Received10/09/2015
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received11/17/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/01/1970
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-