Device Problems
Electrical /Electronic Property Problem (1198); Connection Problem (2900); Protective Measures Problem (3015)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 09/22/2015 |
Event Type
malfunction
|
Manufacturer Narrative
|
Ge healthcare¿s investigation into the reported occurrence is still ongoing.A follow-up report will be issued when the investigation has been completed.Block a: no report of patient involvement.The initial reporter is located outside the u.S., and therefore this information is not provided due to country privacy laws.Device evaluation anticipated, but not yet begun.
|
|
Event Description
|
The hospital reported that, during a preoperative checkout, the anesthesia machine was noted to shut down without an alarm.There was no report of patient involvement.
|
|
Manufacturer Narrative
|
A ge healthcare service representative performed a checkout of the equipment and confirmed the reported complaint.The service representative cleaned and reset all the connections on the control board, resolving the reported complaint.The unit was returned to service.The status of health professional is unknown and was accidentally selected in the initial report; the original reporter occupation is unknown and was accidentally selected in the initial report; the investigation revealed new information regarding the device codes.The above codes are to replace the device codes in the initial report.No further actions are planned at this time.
|
|