TERUMO CARDIOVASCULAR SYSTEMS CORP. CDI BLOOD PARAMETER MONITORING SYSTEM 500; MONITOR, BLOOD-GAS, ON-LINE, CARDIOPULMONARY BYPASS-CDI BPM SYSTEM 500
|
Back to Search Results |
|
Model Number 500AHCT |
Device Problem
Device Operates Differently Than Expected (2913)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 07/16/2015 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).The reported complaint was confirmed.During the laboratory evaluation, power entry module fuses demonstrated continuity and 122 volts alternating current (vac) was observed at the power supply.The power supply output was measured at 12.0 volts direct current (vdc) and met specification.The product surveillance technician (pst) observed the battery switchover integrated circuits (i/c) at u29 on the auxilliary (aux) printed circuit board (pcb) was melted and the capacitor at c81 was blown.All monitor parts were installed on the platter and worked as expected, the aux board was put on separately and the platter would not function.The product will be sent to service to be brought to manufacturers specifications before being returned to the customer.No additional action will be taken at this time.
|
|
Event Description
|
It was reported that during set-up of the device for a cardiopulmonary bypass procedure, the blood parameter monitor (bpm) would not boot-up and the green indicator light was on.As a result, an alternate device was employed.The surgical procedure was completed successfully.There was no delay, no blood loss, nor adverse consequences to the patient.
|
|