• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO CARDIOVASCULAR SYSTEMS CORP. CDI BLOOD PARAMETER MONITORING SYSTEM 500; MONITOR, BLOOD-GAS, ON-LINE, CARDIOPULMONARY BYPASS-CDI BPM SYSTEM 500

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO CARDIOVASCULAR SYSTEMS CORP. CDI BLOOD PARAMETER MONITORING SYSTEM 500; MONITOR, BLOOD-GAS, ON-LINE, CARDIOPULMONARY BYPASS-CDI BPM SYSTEM 500 Back to Search Results
Model Number 500AHCT
Device Problem Device Operates Differently Than Expected (2913)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 07/16/2015
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The reported complaint was confirmed.During the laboratory evaluation, power entry module fuses demonstrated continuity and 122 volts alternating current (vac) was observed at the power supply.The power supply output was measured at 12.0 volts direct current (vdc) and met specification.The product surveillance technician (pst) observed the battery switchover integrated circuits (i/c) at u29 on the auxilliary (aux) printed circuit board (pcb) was melted and the capacitor at c81 was blown.All monitor parts were installed on the platter and worked as expected, the aux board was put on separately and the platter would not function.The product will be sent to service to be brought to manufacturers specifications before being returned to the customer.No additional action will be taken at this time.
 
Event Description
It was reported that during set-up of the device for a cardiopulmonary bypass procedure, the blood parameter monitor (bpm) would not boot-up and the green indicator light was on.As a result, an alternate device was employed.The surgical procedure was completed successfully.There was no delay, no blood loss, nor adverse consequences to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CDI BLOOD PARAMETER MONITORING SYSTEM 500
Type of Device
MONITOR, BLOOD-GAS, ON-LINE, CARDIOPULMONARY BYPASS-CDI BPM SYSTEM 500
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORP.
6200 jackson road
ann arbor MI 48103
Manufacturer Contact
jan winder
6200 jackson road
ann arbor, MI 48103
7346634145
MDR Report Key5141435
MDR Text Key28266241
Report Number1828100-2015-00872
Device Sequence Number1
Product Code DRY
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K133658
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Other
Report Date 10/09/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number500AHCT
Device Catalogue Number500AHCT
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/25/2015
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received Not provided
Initial Date FDA Received10/09/2015
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-