• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS, INC. 1818910 PINN STRAIGHT CUP IMPACTOR; HIPS INSTRUMENTS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS, INC. 1818910 PINN STRAIGHT CUP IMPACTOR; HIPS INSTRUMENTS Back to Search Results
Catalog Number 221750041
Device Problem Peeled/Delaminated (1454)
Patient Problem No Information (3190)
Event Date 09/29/2015
Event Type  malfunction  
Manufacturer Narrative
This complaint is still under investigation.Depuy will notify the fda of the results of this investigation once it has been completed.(b)(4).
 
Event Description
The cup impactor threads are stripped.
 
Manufacturer Narrative
Examination of the returned instrument confirms damage to the threaded tip.Based on the evidence present the root cause is attributed to heavy use / use error.The impactor appears to have been used when not threaded into the mating cup as intended.Based on the investigation, the need for corrective action is not indicated.Continue to monitor via sep-419.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PINN STRAIGHT CUP IMPACTOR
Type of Device
HIPS INSTRUMENTS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic drive
warsaw IN 46582
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic drive
warsaw IN 46582
Manufacturer Contact
chad gibson
700 orthopaedic drive
warsaw, IN 46581
5743725905
MDR Report Key5144120
MDR Text Key28451794
Report Number1818910-2015-32564
Device Sequence Number1
Product Code MRA
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
PP070026
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,health profession
Reporter Occupation Physician
Type of Report Followup
Report Date 09/29/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue Number221750041
Device Lot NumberNT1008
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/05/2015
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received Not provided
Initial Date FDA Received10/12/2015
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received11/30/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/15/2008
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-