• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON ENDO-SURGERY, LLC. ENDOPATH XCEL BLADELESS TROCAR; LAPAROSCOPE, GENERAL AND PLASTIC SURGERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON ENDO-SURGERY, LLC. ENDOPATH XCEL BLADELESS TROCAR; LAPAROSCOPE, GENERAL AND PLASTIC SURGERY Back to Search Results
Catalog Number B12XT
Device Problem Dull, Blunt (2407)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 09/11/2015
Event Type  malfunction  
Manufacturer Narrative
(b)(4).When additional information is received and/or the device analysis has been completed, a supplemental medwatch will be sent.
 
Event Description
It was reported that during a gastric revision procedure, the trocar was not sharp enough to go through fascia.Another like device was used to complete the procedure.There were no adverse consequences for the patient.
 
Manufacturer Narrative
(b)(4).The analysis results found that the b12xt device was returned in good condition.The device was visually inspected and no anomalies were noted with the tip of the obturator and the sleeve.No conclusion could be reached as to what may have caused the reported incident.This report is not intended to deny that you experienced a problem with the device.There may have been other circumstances or issues that occurred during the use of the device that we were unable to duplicate during our laboratory analysis.The batch record was reviewed and no anomalies were noted during the manufacturing process.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOPATH XCEL BLADELESS TROCAR
Type of Device
LAPAROSCOPE, GENERAL AND PLASTIC SURGERY
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo PR 00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC
475 calle c
guaynabo PR 00969
Manufacturer Contact
milton garrett
4545 creek road ml 120a
cincinnati, OH 45242
5133378865
MDR Report Key5144782
MDR Text Key28482031
Report Number3005075853-2015-06460
Device Sequence Number1
Product Code GCJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K032676
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Other
Type of Report Followup
Report Date 09/21/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/05/2019
Device Catalogue NumberB12XT
Device Lot NumberL4F48W
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/23/2015
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received Not provided
Initial Date FDA Received10/13/2015
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received10/29/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/05/2014
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
-
-