• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: INVACARE TAYLOR STREET TDXSP ELECTRIC RED EURO WIDE BASE W/ 23-1 SSD GEAR BOX; WHEELCHAIR, POWERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INVACARE TAYLOR STREET TDXSP ELECTRIC RED EURO WIDE BASE W/ 23-1 SSD GEAR BOX; WHEELCHAIR, POWERED Back to Search Results
Model Number TDXSPREE-CG
Device Problem Sticking (1597)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
Should additional information become available a supplemental record will be filed.  .
 
Event Description
Tbm states that both front locking gas cylinders get hung up in the air when going over a threshold.
 
Manufacturer Narrative
Additional/updated information was added to reflect the device being returned to the manufacturer for evaluation.The original complaint issue was confirmed.The result of the evaluation was that the wheelchair was not configured correctly.A joystick was not returned with the chair; however, when a joystick was added and configured correctly, the chair was functional.Additionally, error codes indicated that the controller had been replaced at some point.
 
Event Description
Tbm states that both front locking gas cylinders get hung up in the air when going over a threshold.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TDXSP ELECTRIC RED EURO WIDE BASE W/ 23-1 SSD GEAR BOX
Type of Device
WHEELCHAIR, POWERED
Manufacturer (Section D)
INVACARE TAYLOR STREET
1200 taylor street
elyria OH 44036
Manufacturer (Section G)
INVACARE TAYLOR STREET
1200 taylor street
elyria OH 44036
Manufacturer Contact
kevin guyton
one invacare way
elyria, OH 44035
8003336900
MDR Report Key5146745
MDR Text Key28634304
Report Number1525712-2015-04841
Device Sequence Number1
Product Code ITI
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor
Reporter Occupation Medical Equipment Company Technician/Representative
Remedial Action Replace
Type of Report Followup
Report Date 12/21/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberTDXSPREE-CG
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/21/2015
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received Not provided
Initial Date FDA Received10/14/2015
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received12/31/2015
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-