• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LIMACORPORATE S.P.A SMR INSERTER-EXTRACTOR HANDLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LIMACORPORATE S.P.A SMR INSERTER-EXTRACTOR HANDLE Back to Search Results
Model Number 9013.02.303
Device Problem Connection Problem (2900)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 07/29/2015
Event Type  malfunction  
Manufacturer Narrative
No anomalies which could have contributed to the issue by checking the manufacturing charts of the lot number of instrument involved.No other similar complaints registered on these lot number, on a total of (b)(6) (handles) and (b)(6) (sleeves) instruments manufactured with these lot number.We will receive the instruments involved and analyze them before submitting our final report.
 
Event Description
Intra-op issue during a surgery dated (b)(6) 2015.It was difficult to insert the sleeve (product code 9013.52.141) onto the handle (product code 9013.02.303).It appeared that the problem was in the handle.The event occurred in (b)(6).
 
Manufacturer Narrative
No anomalies which could have contributed to the issue by checking the manufacturing charts of the lot # of handle involved.No other similar complaints registered on this lot #, on a total of (b)(4) (handles) manufactured with this lot #.We received the handle involved.A dimensional check was performed on the handle and no anomaly was detected on the part of the instrument intended to be connected to the sleeve.The functionality of the handle was tested with a sleeve available in house and the coupling handle-sleeve was functional.A total of three (3) similar complaints were reported, involving difficulty to engage/disengage the sleeve to the handle, on a total of (b)(4) smr inserter-extractor handle manufactured with the product codes 9013.02.302-9013.02.303.The other two cases are still under evaluation (instruments involved not returned yet).No corrective action has been planned for this specific event.Limacorporate will keep monitoring the market on the reoccurrence of this event.
 
Event Description
Intra-op issue during a surgery dated (b)(6) 2015.It was difficult to insert the sleeve (product code 9013.52.141) onto the handle (product code 9013.02.303).It appeared that the problem was in the handle.Event occured in (b)(6).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SMR INSERTER-EXTRACTOR HANDLE
Type of Device
SMR INSERTER-EXTRACTOR HANDLE
Manufacturer (Section D)
LIMACORPORATE S.P.A
via nazionale 52
villanova di san daniele, italy (udine) 33038
IT  33038
Manufacturer (Section G)
LIMACORPORATE S.P.A
via nazionale 52
villanova di san daniele, italy (udine) 33038
IT   33038
Manufacturer Contact
giulio puppa
via nazionale 52
villanova di san daniele, udine 33038
IT   33038
MDR Report Key5147292
MDR Text Key28612799
Report Number3008021110-2015-00041
Device Sequence Number1
Product Code KWT
Combination Product (y/n)N
Reporter Country CodeAS
PMA/PMN Number
K100858
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other
Reporter Occupation Other
Type of Report Followup
Report Date 09/18/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number9013.02.303
Device Lot Number14AA538
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received Not provided
Initial Date FDA Received10/14/2015
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received10/11/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/28/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Hospitalization;
-
-