Model Number SC-2316-50 |
Device Problems
Bent (1059); Break (1069); Fracture (1260); High impedance (1291); Kinked (1339); Failure to Align (2522)
|
Patient Problems
Undesired Nerve Stimulation (1980); Inadequate Pain Relief (2388)
|
Event Date 02/25/2015 |
Event Type
Injury
|
Manufacturer Narrative
|
Additional suspect medical device components involved in the event: model #: sc-2316-50 serial #: (b)(4) description: infinion1x16 perc lead kit-50 cm model #: sc-3400-30 serial #: (b)(4) description: infinion splitter 2x8 kit (30 cm) model #: sc-4316 lot #: 16552647 description: next generation anchor kit-sterile.
|
|
Event Description
|
A report was received that the patient was experiencing uncomfortable stimulation then eventually lost coverage.The patient's lead revealed high impedances on almost all contacts.The physician asked if it was due to the splitters or the clik anchors or both.During the procedure, the physician first tried to disconnect and reconnect splitters, but the result remained the same.The physicians chose to explant everything and replaced both leads and were sutured directly with no clik anchors.The patient was reportedly doing well postoperatively.
|
|
Manufacturer Narrative
|
Additional information was received that another lead splitter was also explanted during the revision procedure.Additional suspect medical device component involved in the event: model #: sc-3400-30, serial/lot #: (b)(4), description: infinion splitter 2x8 kit (30 cm).Sc-2316-50, (sn (b)(4)): device evaluation indicated that the complaint of high lead impedance was confirmed.Visual inspection found proximal contact #16 of the lead was crushed by the splitter setscrew.Additionally, visual inspection found that the lead proximal end was fractured between contacts # 6 and 7.X-ray inspection of the lead revealed that seven cables were completely broken at the bent/kinked location of the lead.No cables were exposed.Because of this all contacts were misaligned with the spring contact.This was the root cause of the high impedance reading.Sc-2316-50 (sn (b)(4)): device evaluation indicated that the complaint had been confirmed.Visual (microscope) and x-ray inspection of the lead revealed that all of the cables were completely broken at the bent/kinked location of the lead, approximately 24 cm from the distal e nd.The bent/kinked location was 1 cm from the set screw mark of the clik anchor.The cables were exposed at the fracture site.The broken cables resulted in the reported complaint of high impedance.Sc-3400-30 (sn (b)(4)): device evaluation indicated that the devices passed all tests performed.The complaint was not verified.A test infinion lead was inserted into the splitters connector without any anomalies.Devices exhibited normal device characteristics.Sc-4316 (lot# 16552647): device evaluation indicated that visual inspection found that one of the clik anchors had torn eyelets, and a small piece of silicone was not returned.The other clik anchor exhibited normal characteristics.
|
|
Event Description
|
A report was received that the patient was experiencing uncomfortable stimulation then eventually lost coverage.The patient's lead revealed high impedances on almost all contacts.The physician asked if it was due to the splitters or the clik anchors or both.During the procedure, the physician first tried to disconnect and reconnect splitters, but the result remained the same.The physicians chose to explant everything and replaced both leads and were sutured directly with no clik anchors.The patient was reportedly doing well postoperatively.
|
|
Manufacturer Narrative
|
Additional information was received that the missing silicone was removed from the patient¿s body.The physician confirmed that all products were removed.
|
|
Event Description
|
A report was received that the patient was experiencing uncomfortable stimulation then eventually lost coverage.The patient's lead revealed high impedances on almost all contacts.The physician asked if it was due to the splitters or the clik anchors or both.During the procedure, the physician first tried to disconnect and reconnect splitters, but the result remained the same.The physicians chose to explant everything and replaced both leads and were sutured directly with no clik anchors.The patient was reportedly doing well postoperatively.
|
|
Search Alerts/Recalls
|