• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CAREFUSION VELA VENTILATOR; VENTILATOR, CONTINUOUS, FACILITY USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CAREFUSION VELA VENTILATOR; VENTILATOR, CONTINUOUS, FACILITY USE Back to Search Results
Model Number VELA COMP D
Device Problems Image Display Error/Artifact (1304); Delamination (2904)
Patient Problem No Patient Involvement (2645)
Event Date 09/14/2015
Event Type  malfunction  
Manufacturer Narrative
Carefusion file identification number is (b)(4).Any additional information that is provided by the customer will be included in a follow-up report.The device has not been received by carefusion.
 
Event Description
The customer reported while using the vela ventilator the front panel became non-responsive secondary to display panel delamination.The customer reported no patient involvement or allegation of patient harm.At this time, carefusion has not received the suspect device component for evaluation.
 
Manufacturer Narrative
Per results of investigation, the carefusion failure analysis (fa) lab received the suspect front panel assembly and installed in on a known good vela ventilator unit.The panel was inspected and found "liquid ingress" on it.The unit was powered on and the reported complaint of the front panel becoming non-responsive due to delamination was duplicated.This issue will be internally investigated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VELA VENTILATOR
Type of Device
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
CAREFUSION
22745 savi ranch parkway
yorba linda CA 92887
Manufacturer (Section G)
CAREFUSION
1100 bird center drive
palm springs CA 92262
Manufacturer Contact
jill rittorno
75 n. fairway drive
vernon hills, IL 60061
2627587805
MDR Report Key5148666
MDR Text Key28574490
Report Number2021710-2015-01886
Device Sequence Number1
Product Code CBK
Combination Product (y/n)N
Reporter Country CodeSF
PMA/PMN Number
K093094
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other Health Care Professional
Type of Report Followup
Report Date 09/16/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberVELA COMP D
Device Catalogue Number16532-02
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received Not provided
Initial Date FDA Received10/14/2015
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received05/04/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/30/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-