MAUDE Adverse Event Report: CYBERONICS - HOUSTON PROGRAMMING SOFTWARE; PROGRAMMING COMPUTER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CYBERONICS - HOUSTON PROGRAMMING SOFTWARE; PROGRAMMING COMPUTER Back to Search Results
Model Number MODEL 250
Device Problems Failure to Interrogate (1332); Failure to Advance (2524)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/18/2015
Event Type  malfunction  
Event Description
It was reported that the programming wand was malfunctioning and was unable to perform interrogations.The 9v batteries had been previously replaced in both programming wands prior to the reported event and the green light was confirmed to stay on for longer than 25 seconds.The handheld device was unplugged when performing the interrogations and the connections between the programming system components were secure.The interrogations only occur in two rooms in the clinic and interrogations in these rooms have been previous successful; therefore, emi was not suspected to be the issue.The programming system was returned to the manufacturer for analysis.The programming wand and software flashcard performed according to functional specifications.During the analysis, it was identified that the handheld was unable to advance past the screen alignment utility.The cause for the display anomaly is associated with resistance values being higher than expected in the touch screen circuitry.No further anomalies associated with the handheld performance were identified during the analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROGRAMMING SOFTWARE
Type of Device
PROGRAMMING COMPUTER
Manufacturer (Section D)
CYBERONICS - HOUSTON
100 cyberonics blvd
houston TX 77058
Manufacturer (Section G)
CYBERONICS - HOUSTON
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
njemile crawley
100 cyberonics blvd
suite 600
houston, TX 77058
2812287200
MDR Report Key5151299
MDR Text Key28707237
Report Number1644487-2015-06126
Device Sequence Number1
Product Code LYJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other
Report Date 10/07/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMODEL 250
Device Lot Number1063824
Other Device ID NumberVERSION 8.1
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/03/2015
Is the Reporter a Health Professional? No
Event Location Other
Initial Date Manufacturer Received Not provided
Initial Date FDA Received10/15/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/30/2012
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-