A steris service technician arrived at the user facility and was able to duplicate the reported event.The technician was informed the surgical floor may have been wet at the time of the event creating a suction effect on the floor lock foot pads.The technician met with the facility's or manager to discuss proper cleaning techniques for the 5085 surgical table and recommended the facility ensure the floor is completely dry before locking the feet to the floor.The technician replaced the floor lock foot pads, tested the table and returned the table to service.No further issues have been reported.
|