U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ROCHE DIAGNOSTICS ROCHE ELECSYS ANTI-CCP IMMUNOASSAY; ANTIBODIES, ANTI-CYCLIC CITRULLINATED PEPTIDE (CCP)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ROCHE DIAGNOSTICS ROCHE ELECSYS ANTI-CCP IMMUNOASSAY; ANTIBODIES, ANTI-CYCLIC CITRULLINATED PEPTIDE (CCP) Back to Search Results
Catalog Number 05031656190
Device Problem High Test Results (2457)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 06/08/2015
Event Type  malfunction  
Manufacturer Narrative
This event occurred in (b)(6).
 
Event Description
The customer received questionable anti-ccp results for one patient from an unknown roche analyzer.On (b)(6) 2015, the result was 478 u/ml, but the result by stacia, meblux test whose principle was cleia, was negative.Additionally, the patient's rf was negative.It was unclear if this testing was performed with the same sample.On (b)(6) 2015, the result was 453 u/ml, but the result for the same sample with by stacia, meblux test was negative.The rf result was negative and the anti-nuclear antibody (ana) was <40 folds.Information concerning if any erroneous result was reported outside the laboratory was requested, but was not provided.The patient was not adversely affected.
 
Manufacturer Narrative
A specific root cause could not be identified as the sample was not available for further investigation.A possible cause for the issue could have been an interference in the sample.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ROCHE ELECSYS ANTI-CCP IMMUNOASSAY
Type of Device
ANTIBODIES, ANTI-CYCLIC CITRULLINATED PEPTIDE (CCP)
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhoferstrasse 116
na
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
michael leslie
9115 hague road
na
indianapolis, IN 46250
3175214343
MDR Report Key5162018
MDR Text Key28805983
Report Number1823260-2015-04357
Device Sequence Number1
Product Code NHX
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K081338
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Reporter Occupation Health Professional
Type of Report Followup
Report Date 12/16/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number05031656190
Device Lot NumberASKU
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Event Location Laboratory
Initial Date Manufacturer Received Not provided
Initial Date FDA Received10/20/2015
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received12/16/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-