Model Number SC-2218-50 |
Device Problems
Bent (1059); Fracture (1260); High impedance (1291); Kinked (1339)
|
Patient Problem
Inadequate Pain Relief (2388)
|
Event Date 08/20/2015 |
Event Type
Injury
|
Manufacturer Narrative
|
Additional suspect medical device components involved in the event: model #: sc-2218-50 serial #: (b)(4) description: linear st lead, 50cm model #: sc-2218-70 serial #: (b)(4) description: linear st lead, 70cm.
|
|
Event Description
|
A report was received that the patient was experiencing loss of stimulation.High impedances were noted on leads.The patient will undergo lead replacement procedure.
|
|
Manufacturer Narrative
|
Additional information was received that database analysis showed that the impedance measurements revealed high values on several contacts on the leads and were not functioning.
|
|
Event Description
|
A report was received that the patient was experiencing loss of stimulation.High impedances were noted on leads.The patient will undergo lead replacement procedure.
|
|
Manufacturer Narrative
|
Additional information was received that the patient underwent a revision procedure wherein two leads were replaced due to suspected fracture.The patient was reportedly doing well post operatively.
|
|
Event Description
|
A report was received that the patient was experiencing loss of stimulation.High impedances were noted on leads.The patient will undergo lead replacement procedure.
|
|
Manufacturer Narrative
|
Sc-2218-50 (sn (b)(4)) the complaint has been confirmed.Visual (microscope) and x-ray inspection of the lead revealed that all of the cables were completely broken at the bent/kinked location of the lead.The bent/kinked locations are 1 cm from both sides of the set screw mark of the clik anchor.There are no exposed cables at the fracture locations.Sc-2218-50 (sn (b)(4)) the complaint has been confirmed.Visual (microscope) and x-ray inspection of the lead revealed that 2 cables were completely broken at the bent/kinked location of the lead.The bent/kinked location is 1 cm from the set screw mark of the clik anchor.There are no exposed cables at the fracture locations.
|
|
Event Description
|
A report was received that the patient was experiencing loss of stimulation.High impedances were noted on leads.The patient will undergo lead replacement procedure.
|
|
Event Description
|
A report was received that the patient was experiencing loss of stimulation.High impedances were noted on leads.The patient will undergo lead replacement procedure.
|
|
Search Alerts/Recalls
|