• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAQUET CARDIOPULMONARY AG MAQUET HCU40 DEVICE; CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAQUET CARDIOPULMONARY AG MAQUET HCU40 DEVICE; CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 70104.4054
Device Problem Infusion or Flow Problem (2964)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/21/2015
Event Type  Injury  
Manufacturer Narrative
October 21, 2015 12:03 pm (gmt-4:00) added by (b)(6): (b)(4).A maquet field service technician was on site and investigated the unit in question.The technician checked the cardioplegia circuit and the ice formation.After troubleshooting the device, the technician could confirm the reported issue with the low flow because the ice block forming setup on the cardioplegia side was set too high.The setup for the ice formation was set to 2/3 which caused the problem with the low flow.The technician cleaned the unit and changed the water.Functionality tests were performed successfully.A supplemental medwatch will be submitted as soon as additional information becomes available.
 
Event Description
It was reported that the flow on the cardioplegia side of a hcu40 device is too low.The device was exchanged immediately.No patient involvement (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAQUET HCU40 DEVICE
Type of Device
CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
MAQUET CARDIOPULMONARY AG
rastatt
GM 
Manufacturer (Section G)
MICHAEL CAMPBELL
maquet cardiopulmonary ag
kehler strasse 31
rastatt 76437
GM   76437
Manufacturer Contact
maquet cardiopulmonary ag
kehler strasse 31
rastatt 76437
MDR Report Key5166820
MDR Text Key28906052
Report Number8010762-2015-01119
Device Sequence Number1
Product Code DWC
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K130300
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,foreign,health profe
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 09/25/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number70104.4054
Device Catalogue Number70104.4054
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/25/2015
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 09/25/2015
Initial Date FDA Received10/21/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/01/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-