Sorin group (b)(4) manufactures the cp5 flow module.The incident occurred in (b)(6).This medwatch report is being filed on behalf of sorin group (b)(4).Sorin group received a report that the cp5 flow module displayed values incorrectly.There was no patient involvement.A sorin group field service representative was dispatched to the facility.The flow module was replaced and returned to sorin group (b)(4) for further investigation.The issue was reproduced after a 30 minute test run.A hardware analysis was conducted, which revealed a faulty pcb board.The faulty board was replaced and a test run of 24 hours was conducted.No further issues were identified.Due to the age of the board, it was determined that no further investigation is required, and the faulty board was discarded.A review of the dhr could not identify any concessions, deviations or nonconformities relevant to the reported failure.This issue will be monitored for trends and if a trend is identified, corrective action will be recommended.
|