• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NIHON KOHDEN TOMIOKA CORPORATION AG-920RA; MULTI-GAS UNIT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NIHON KOHDEN TOMIOKA CORPORATION AG-920RA; MULTI-GAS UNIT Back to Search Results
Model Number AG-920RA
Device Problem Device Inoperable (1663)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/22/2015
Event Type  malfunction  
Manufacturer Narrative
The customer reported that the multi-gas unit is giving a calibration error.The o2 reading is coming in at half the value it should be.The biomedical engineer has replaced the dryline and the issue still persists.The device was returned to nihon (b)(4), evaluated, and the calibration error was duplicated.The device was disassembled and the internal gas was replaced to fix this issues.Gases are in specifications in regards to the service/operators manual.All functions were tested prior to shipment.The device was returned to the customer.Nihon (b)(4) will submit a supplemental report in accordance with 21 cfr part 803.56 if additional information becomes available.
 
Event Description
The customer reported that the multi-gas unit is giving a calibration error.The o2 reading is coming in at half the value it should be.The biomedical engineer has replaced the dryline and the issue still persists.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AG-920RA
Type of Device
MULTI-GAS UNIT
Manufacturer (Section D)
NIHON KOHDEN TOMIOKA CORPORATION
486 nanokaichi
tomioka city, japan 370-2 343
JA  370-2343
Manufacturer (Section G)
NIHON KOHDEN TOMIOKA CORPORATION
486 nanokaichi
tomioka city, japan 370-2 343
JA   370-2343
Manufacturer Contact
thomas bento
1-31-4 nishiochia, shinjuku-ku
tokyo, japan 161-8-560
JA   161-8560
2687708
MDR Report Key5171766
MDR Text Key29193697
Report Number8030229-2015-00317
Device Sequence Number1
Product Code CCK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K020046
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 10/22/2015,09/22/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberAG-920RA
Device Catalogue NumberAG-920RA
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/30/2015
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Report Sent to FDA10/22/2015
Distributor Facility Aware Date09/22/2015
Device Age50 MO
Event Location Hospital
Date Report to Manufacturer10/22/2015
Initial Date Manufacturer Received 10/22/2015
Initial Date FDA Received10/23/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/22/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-