The customer reported that the multi-gas unit is giving a calibration error.The o2 reading is coming in at half the value it should be.The biomedical engineer has replaced the dryline and the issue still persists.The device was returned to nihon (b)(4), evaluated, and the calibration error was duplicated.The device was disassembled and the internal gas was replaced to fix this issues.Gases are in specifications in regards to the service/operators manual.All functions were tested prior to shipment.The device was returned to the customer.Nihon (b)(4) will submit a supplemental report in accordance with 21 cfr part 803.56 if additional information becomes available.
|