ACCLARENT, INC. RELIEVA SPIN BALLOON SINUPLASTY DEVICE; INSTRUMENT, ENT MANUAL SURGICAL Back to Search Results
Model Number N/A
Device Problems Detachment Of Device Component (1104); Component Falling (1105); Material Rupture (1546)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/25/2015
Event Type  Injury  
Manufacturer Narrative
The subject device is pending return for evaluation.Review of manufacturing records associated with the subject device did not detect any anomalies.A supplemental report will be submitted when the product is returned and tested.Acclarent will continue to monitor this phenomenon for trending purposes.
 
Event Description
Acclarent was made aware of an event that occurred during a procedure in which an acclarent relieva spin balloon sinuplasty system was used.During the primary bilateral maxillary balloon sinuplasty procedure (bsp) along with septoplasty and turbinate surgery, there was no difficulty reported in accessing the patient's sinus cavities.Upon the first inflation of the acclarent relievapin balloon, the balloon was said to have ruptured and become detached, falling into the patient's maxillary sinus cavity.The surgeon performed an antrostomy to retrieve the balloon, and confirmed that all pieces were successfully removed.There was no other intervention required, and the procedure was completed using another acclarent relievapin balloon sinuplasty system.The patient was reported to have been doing well with no sequelae reported following the procedure.
 
Manufacturer Narrative
Acclarent received the package on 2/24/2016.The returned package included the following items: a spin sinuplasty balloon, luma sentry wire, balloon slider, wire slider, detachable connector, guide catheter (m-110).Upon receiving the package, before decontamination, the product was visually inspected and it was observed that the tip of the guide catheter was kinked, part of the balloon was missing, the tip of balloon catheter was damaged and inner shaft of balloon catheter was kinked.No product analysis could be performed due part of the balloon being missing.The customer complaint has been verified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RELIEVA SPIN BALLOON SINUPLASTY DEVICE
Type of Device
INSTRUMENT, ENT MANUAL SURGICAL
Manufacturer (Section D)
ACCLARENT, INC.
1525-b o'brien dr
menlo park CA
Manufacturer Contact
izabel nielson
1525-b o'brien dr
menlo park, CA 94025
6506877492
MDR Report Key5174910
MDR Text Key29212010
Report Number3005172759-2015-00015
Device Sequence Number1
Product Code LRC
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K111875
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date 09/26/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/11/2018
Device Model NumberN/A
Device Catalogue NumberRS0616MFS
Device Lot Number150811C-CM
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received Not provided
Initial Date FDA Received10/24/2015
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received03/11/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/19/2015
Is the Device Single Use? Yes
Type of Device Usage N
Removal/Correction NumberN/A
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-