• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNTHES SELZACH INSERTION HANDLE FOR SUPRAPATELLAR; NAIL, FIXATION, BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES SELZACH INSERTION HANDLE FOR SUPRAPATELLAR; NAIL, FIXATION, BONE Back to Search Results
Catalog Number 03.010.440
Device Problems Difficult to Remove (1528); Sticking (1597); Separation Failure (2547)
Patient Problem Sedation (2368)
Event Date 10/12/2015
Event Type  Injury  
Manufacturer Narrative
Patient information is not available for reporting.Device is an instrument and is not implanted or explanted.(b)(4).Subject device has been received; no conclusions could be drawn as the device is entering the complaint system.Device history record review: manufacturing location: (b)(4) - manufacturing date: october 1, 2014.Non-conformance reports (ncr) were generated during the manufacturing of the articles.One article was prepared without labeling and the package was defective.One part was dropped on the floor.A critical defect for defective packaging was also addressed.All the parts were booked in quarantine storage.Device used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
Device report from synthes (b)(4) reports an event in (b)(6) as follows: it was reported that a connection screw could not be separated from an insertion handle during a surgical procedure on (b)(6), 2015.Per the reporter, several attempts were made with all of the appropriate instruments, but the devices could not be separated from one another.It was necessary for the surgeon to remove the nail and replace it with another.The procedure was prolonged by sixty (60) minutes or more.This report is 2 of 3 for (b)(4).
 
Manufacturer Narrative
Device was used for treatment, not diagnosis.Synthes lot number was added.The previously report lot number, 14-4638, is the supplier lot number.A manufacturing investigation was performed for the subject device.The subject device was returned along with the complained nail and connecting screw.A functional test of the received articles showed that the devices could be mounted together without any problems.The complaint issue could not be replicated.Unfortunately without more information or x-rays we are not able to determine the root cause which has led to this complaint.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INSERTION HANDLE FOR SUPRAPATELLAR
Type of Device
NAIL, FIXATION, BONE
Manufacturer (Section D)
SYNTHES SELZACH
bohnackerweg 5
selzach CH254 5
SZ  CH2545
Manufacturer (Section G)
SYNTHES SELZACH
bohnackerweg 5
selzach CH254 5
SZ   CH2545
Manufacturer Contact
linda plews
1302 wrights lane east
west chester, PA 19380
6107195000
MDR Report Key5180521
MDR Text Key29358527
Report Number3000270450-2015-10218
Device Sequence Number1
Product Code JDS
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
PK111667
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 10/12/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number03.010.440
Device Lot NumberU208643
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/20/2015
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/12/2015
Initial Date FDA Received10/27/2015
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received12/02/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/01/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age35 YR
-
-