Model Number XT245HSJ
Device Problem Leak/Splash (1354)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 09/07/2015
Event Type  malfunction  
Manufacturer Narrative
One double lumen pediasat catheter was returned for evaluation.As received, the catheter tube was cut and the distal side of the catheter was not returned.Dry blood was observed around the optic module connector.Leak testing confirmed an interlumen leakage in the backform between the distal and optic lumens.No leakage or occlusion was observed in the proximal lumen.Further examination found the catheter to be cut (clean sharp edges) 2.5cm distal of the backform.The distal section of the catheter tip was not returned.An x-ray of the backform was taken which confirmed a void inside the backform.Visual examinations were performed under microscope at 10x magnification.The lot number was not provided; therefore, a review of the manufacturing records could not be completed.Customer report of leakage issue was confirmed.An investigation has been initiated to consider any potential manufacturing factors that may have contributed to this complaint and implement any necessary corrective actions.
 
Event Description
It was reported that blood leakage was observed from the extension tube during use.There were no patient complications reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRESEP CENTRAL VENOUS OXIMETRY SET
Type of Device
PRESEP CATHETER
Manufacturer (Section D)
EDWARDS LIFESCIENCES, PR
state rd indus pk 402 km 1.4
anasco PR 00610
Manufacturer (Section G)
EDWARDS LIFESCIENCES, PR
state rd indus pk 402 km 1.4
anasco PR 00610
Manufacturer Contact
lynn selawski
1 edwards way
irvine, CA 92614
9492504386
MDR Report Key5181954
MDR Text Key29407982
Report Number2015691-2015-02883
Device Sequence Number1
Product Code DQE
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K053609
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial
Report Date 09/07/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberXT245HSJ
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/15/2015
Initial Date Manufacturer Received 09/07/2015
Initial Date FDA Received10/27/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-