Boston scientific received information that this patient stated he had not been feeling well and was experiencing lightheadedness and his heart rate was 60 to 65 beats per minute.In addition, the physician stated that the patient needed a device change out.A boston scientific technical services consultant advised the patient to contact the clinic about the experienced symptoms.The device was explanted four months after this patient's call.Physical damage to the device header was observed.Additionally, the lead was stuck in the device and setscrew issues were experienced.No adverse patient effects were reported.
|
Upon receipt in our post market quality assurance laboratory, a thorough evaluation of the device was performed.Visual inspection noted medical adhesive in the lead barrel which could cause the lead to become stuck.The device was then exposed to simulated heart load conditions, and the pacing, and sensing functions were tested.Impedance testing was completed and all measurements were within normal limits.The device operated appropriately with no interruptions in therapy output at the returned programmed settings.A series of electrical tests was also performed, and again, normal device function was observed.
|