(b)(4).Age at time of event: 18 years or older.(b)(4).Device evaluated by manufacturer - the device has a kink at 13mm from the tip.In addition has char in the base of the tip and the ring #1 and broken adhesive on it.The ablation was verified by using the maestro generator 4000 and the device was found within specifications.Electrical test was performed and the device was found out of specifications.The device has a intermittent short between tip and r1.The handle was opened finding no issues to the connector.The stiffening rod was removed identified a possible affected area, however, no issues were identified.The stiffening rod has not damages on it.The device was dissected after the torque hole finding the following: the r1 wire is peeled close to the ring soldering.This area match with the kink in the center support and a burning mark in the tip wire.This caused the short between tip and r1.Evidence of fluids was found inside the shaft.Regarding the kink in the distal section the device was dissected finding the center support kinked at 14mm from the tip.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The most probable root cause is operational context as device performance was limited due to anatomical procedural factors.(b)(4).
|