Upon visual inspection, the complaint has been confirmed that the tip is broken.The ifu for this part states that "the tip of the instrument is extremely thin and delicate, care should be taken to avoid applying significant pressure to the tip." the device history was reviewed and no non-conformance was found for this lot.There are no indication of manufacturing defects.The most likely underlying cause is excessive force.Supplemental report two of two for the same event, see also 0001032347-2015-00427-1.
|