• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOMET MICROFIXATION ELEVATOR #301

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOMET MICROFIXATION ELEVATOR #301 Back to Search Results
Model Number N/A
Device Problem Break (1069)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
Current information is insufficient to permit a valid conclusion about the cause of this event.A follow up report will be sent upon completion of the device evaluation.
 
Event Description
The facility reported an instrument broke during a procedure.The facility confirmed parts were retrieved at the time of the procedure.No adverse events were reported.
 
Manufacturer Narrative
One (1) elevator #301 (pn 09-0257, lot#656220, manufacture date: october 1999) was returned without original packaging for a broken tip.The manufacturing history was reviewed and no non-conformances associated with this lot of parts were identified.The elevator was visually evaluated and a small portion of the tip was found to be fractured.The broken tip fragment was not returned.The elevator shows signs of normal wear from typical extended use.The most likely cause of the complaint was operation of the instrument outside of the intended use, resulting in failure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELEVATOR #301
Type of Device
ELEVATOR
Manufacturer (Section D)
BIOMET MICROFIXATION
1520 tradeport drive
jacksonville FL 32218
Manufacturer (Section G)
BIOMET MICROFIXATION
1520 tradeport drive
jacksonville FL 32218
Manufacturer Contact
michelle cole
1520 tradeport drive
jacksonville, FL 32218
9047414400
MDR Report Key5189171
MDR Text Key30322917
Report Number0001032347-2015-00429
Device Sequence Number1
Product Code EMJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
PEXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 09/30/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number09-0257
Device Lot Number656220
Other Device ID NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/09/2015
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 09/30/2015
Initial Date FDA Received10/29/2015
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received12/03/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/06/1999
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Removal/Correction NumberN/A
Patient Sequence Number1
-
-