Model Number WCD 4000
Device Problem Cut In Material (2454)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/26/2015
Event Type  malfunction  
Manufacturer Narrative
Device evaluation of belt sn (b)(4) has been completed.The reported problem (damaged cable) was confirmed.Upon evaluation, the cable connecting ecg electrodes a and b with the front therapy electrode (te) was pulled from the front te.The root cause of the damaged cable is excessive force placed on the cable.No adverse event resulted from the damaged cable.
 
Event Description
A us distributor contacted zoll to report that a patient's electrode belt cable was cut.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFEVEST WCD 4000 SYSTEM
Type of Device
WEARABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ZOLL MANUFACTURING CORPORATION
121 gamma drive
pittsburgh PA 15238 3495
Manufacturer (Section G)
ZOLL MANUFACTURING CORPORATION
121 gamma drive
pittsburgh PA 15238 3495
Manufacturer Contact
katelynn mains
121 gamma drive
pittsburgh, PA 15238-3495
4129683333
MDR Report Key5190947
MDR Text Key30393038
Report Number3008642652-2015-06917
Device Sequence Number1
Product Code MVK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010030
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor
Reporter Occupation Other
Type of Report Initial
Report Date 10/27/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberWCD 4000
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/07/2015
Initial Date Manufacturer Received 10/08/2015
Initial Date FDA Received10/30/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/06/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse