Device evaluation summary: device evaluation of electrode belt sn (b)(4) is complete.The reported problem (damaged cable with wires exposed) was confirmed.As received, the electrode belt failed incoming testing.Upon investigation, the cable connecting the distribution node (dn) to rear therapy pad (te) was pulled from the strain relief.The cause of the test failure was the pulled dn to rear te cable.The root cause of the pulled dn to rear te cable was excessive force.No adverse event resulted from the defective electrode belt.
|