• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY SYNTHES POWER TOOLS 6.1CM ATTACHMENT, BM; BUR, SURGICAL, GENERAL & PLASTIC SURGERY - ATTACHMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY SYNTHES POWER TOOLS 6.1CM ATTACHMENT, BM; BUR, SURGICAL, GENERAL & PLASTIC SURGERY - ATTACHMENT Back to Search Results
Catalog Number B-SILVER_LL
Device Problem Overheating of Device (1437)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 10/14/2015
Event Type  malfunction  
Manufacturer Narrative
This device was returned for service, however did not meet manufacturing specifications during pre-repair assessment.Reliability engineering evaluated the device and observed that the device temperature was over specifications and the bearings were worn.Therefore, the reported condition was confirmed.The assignable root cause was determined to be due to normal use over time.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Event Description
It was reported that during service and repair/pre-testing, it was observed that the attachment device was heating up.The event was not related to surgery.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
6.1CM ATTACHMENT, BM
Type of Device
BUR, SURGICAL, GENERAL & PLASTIC SURGERY - ATTACHMENT
Manufacturer (Section D)
DEPUY SYNTHES POWER TOOLS
4500 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
DEPUY SYNTHES POWER TOOLS
4500 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
rodney crain
4500 riverside drive
palm beach gardens, FL 33410
5616271080
MDR Report Key5191802
MDR Text Key30406921
Report Number1045834-2015-12352
Device Sequence Number1
Product Code GFF
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
PK974025
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Health Professional
Type of Report Initial
Report Date 10/14/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberB-SILVER_LL
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/05/2015
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/14/2015
Initial Date FDA Received10/30/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/17/1997
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-