• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT MEDICAL OPTICS EMERALD; SURGICAL ADJUNCTS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MEDICAL OPTICS EMERALD; SURGICAL ADJUNCTS Back to Search Results
Model Number EMERALDC30
Device Problem Material Deformation (2976)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/02/2015
Event Type  malfunction  
Manufacturer Narrative
(b)(6).All pertinent information available to abbott medical optics has been submitted.
 
Event Description
It was reported that the surgeon observed the tip part was juggy/burr (tip deformed); therefore, the surgeon did not use it.Reportedly, the operation was completed by using another emeraldc30 cartridge and there was no patient contact with the initial cartridge.No further information was provided.
 
Manufacturer Narrative
The cartridge was returned to the manufacturer for evaluation.Visual inspection of the return sample revealed that the cartridge tip was dent/distortions and scarce amounts of viscoelastic solution were observed on cartridge which indicate that the unit was handled and prepare for surgical use.The reported complaint is confirmed.A review of the directions for use (dfu) was conducted.The dfu adequately provides instructions, precautions, and warning for the proper use and handling of the device.The manufacturing record review cannot be performed due to an unknown lot number.The lot number was not provided by the customer.All pertinent information available to abbott medical optics has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EMERALD
Type of Device
SURGICAL ADJUNCTS
Manufacturer (Section D)
ABBOTT MEDICAL OPTICS
santa ana CA
Manufacturer (Section G)
ABBOTT MEDICAL OPTICS INC.
road 402 north, km 4.2
anasco industrial park, pob 14
anasco PR 00610
Manufacturer Contact
pam mcclain
1700 east st. andrew place
santa ana, CA 92705
7142478243
MDR Report Key5193251
MDR Text Key30422356
Report Number2648035-2015-01084
Device Sequence Number1
Product Code KYB
Combination Product (y/n)N
PMA/PMN Number
K961242
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 12/27/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberEMERALDC30
Device Catalogue NumberEMERALDC30
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/26/2015
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/02/2015
Initial Date FDA Received10/31/2015
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received12/28/2015
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-