• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ANATOMAGE INC. ANATOMAGE GUIDE ; SURGICAL GUIDE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ANATOMAGE INC. ANATOMAGE GUIDE ; SURGICAL GUIDE Back to Search Results
Device Problems Product Quality Problem (1506); Device Dislodged or Dislocated (2923)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/15/2015
Event Type  Injury  
Event Description
The doctor reported that when he put the guide on the pt the fit was slightly tight but fine.Doctor laid a flap at both sites #13 and #14.He drilled the pilot drill, took a pa, and noticed #13 was too mesial and was close to the root of #12.If the doctor had continued with the drilling sequence, he may have hit the root of #12.The doctor reported that he free-handed the rest of the surgery and moved both implants more distal resulting in successful implant placement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANATOMAGE GUIDE
Type of Device
SURGICAL GUIDE
Manufacturer (Section D)
ANATOMAGE INC.
san jose CA
Manufacturer Contact
shirley lee
111 n market st ste 500
san jose, CA 95113
4088851474
MDR Report Key5200627
MDR Text Key30457004
Report Number3008272529-2015-00021
Device Sequence Number1
Product Code NDP
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Dentist
Type of Report Initial
Report Date 09/24/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer10/01/2015
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 09/24/2015
Initial Date FDA Received10/26/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/01/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age69 YR
Patient Weight59
-
-