The pigtail of the catheter cracked along the length of the black part during injection.The procedure was stopped and the device was changed.According to the initial reporter, there were no consequences to the patient.Additional information has been requested, but not provided by the reporter at the time of this report.
|
(b)(4).Investigation - evaluation a review of complaint history, device record history, instructions for use (ifu), quality control, specifications and visual inspection of the returned device was conducted during the investigation.Two devices were returned in an opened and used condition.Visual inspection confirms material failure.One device had circumferential separation 8 mm distal of the bond joint, complete circumferential separation 1.5 mm distal of the bond joint and longitudinal splits running the length of the separated tip.The other device has 3 longitudinal splits beginning 1.0 cm distal the bond joint and extending to the tip.The device is shipped with an instruction for use (ifu) that describes the intended use, specific items are addressed such as: "due to thinwall construction, extreme care must be exercised during manipulation and withdrawal.Catheter insertion through a synthetic vascular graft should be avoided whenever possible." "the possible whiplash effect of the long, soft catheter tip must be considered during selective angiography." a capa request was initiated as part of the recall actions to further investigate the incident.The device is part of recall z-2317/2319-2015 / res: 71692 z-0798-2016 / res72387.The appropriate internal personnel have been notified and we will continue to monitor for similar complaints.
|
The pigtail of the catheter cracked along the length of the black part during injection.The procedure was stopped and the device was changed.According to the initial reporter, there were no consequences to the patient.Additional information has been requested, but not provided by the reporter at the time of this report.
|