Model Number 5001102 |
Device Problems
Leak/Splash (1354); Failure to Prime (1492)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 10/20/2015 |
Event Type
Injury
|
Manufacturer Narrative
|
There are no devices of this lot remaining in inventory for analysis.Quest medical, inc.Has limited information related to the patient's medical history and is unable to form an opinion as to the relevancy of the patient's history to the event reported.Quest medical, inc.Defers to the patient's physician regarding medical history.
|
|
Event Description
|
The hospital perfusionist reported an issue encountered with the mps delivery set during priming.He reported that the delivery set was leaking on the lower half of the heat exchanger.The perfusionist stated it appeared to be a failure in the bonding of the two pieces.As a result, the delivery set was changed out for another prior to use in the surgery.There were no patient complications reported as a result of the alleged event.The device was returned to the manufacturer for analysis.
|
|
Manufacturer Narrative
|
The customer originally reported the lot to be 0494465u07, but upon examination of the returned device along with manufacturing records, the lot was for 0495215u11.The complaint sample analysis found that the uv bond was inadequate.Because heat exchangers are 200% leak-tested and any rejects are automatically sorted from the lot, it is unlikely this device had passed leak testing.The automatic robot function was not working during this time.The rejected devices were being manually sorted by the operators.It is most likely th eoperator mixed this device with the good parts/placed in the incorrect bin.
|
|
Search Alerts/Recalls
|