• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAUSCH + LOMB TRULIGN TORIC POSTERIOR CHAMBER IOL; LENS, INTRAOCULAR, TORIC OPTICS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAUSCH + LOMB TRULIGN TORIC POSTERIOR CHAMBER IOL; LENS, INTRAOCULAR, TORIC OPTICS Back to Search Results
Model Number BL1UT
Device Problem Mechanical Problem (1384)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  Injury  
Manufacturer Narrative
Investigation of this event is in progress.A supplemental report will be submitted upon completion of the investigation.
 
Event Description
It was reported that a lens was explanted a lens due to mechanical issue.Reportedly, the lens was not accommodating properly.Additional information has been requested but has not been received.
 
Manufacturer Narrative
The lens was returned to b+l.All dimensional measurements were found to be within specifications.Lens evaluation did not identify any anomaly with the lens.The device history records were reviewed and there were no discrepancies or unusual finding that relate to the reported issue.Based on the information available, the root cause of the reported event could not be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRULIGN TORIC POSTERIOR CHAMBER IOL
Type of Device
LENS, INTRAOCULAR, TORIC OPTICS
Manufacturer (Section D)
BAUSCH + LOMB
rochester NY 14609
Manufacturer (Section G)
BAUSCH + LOMB
10574 acacia street suite d1
rancho cucamonga CA 91730
Manufacturer Contact
sharon spencer
50 technology drive west
irvine, CA 92618
9493985698
MDR Report Key5208872
MDR Text Key30681065
Report Number0001313525-2015-02937
Device Sequence Number1
Product Code MJP
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030002
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 10/15/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date09/03/2018
Device Model NumberBL1UT
Device Lot Number024873
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/15/2015
Initial Date FDA Received11/06/2015
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received12/30/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/06/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age68 YR
-
-