KARL LEIBINGER MEDIZINTECHNIK GMBH & CO. KG MAXDRIVE MINI SCREW; BONE SCREW Back to Search Results
Model Number 25-872-11-91
Device Problems Break (1069); Detachment of Device or Device Component (2907)
Patient Problem Device Embedded In Tissue or Plaque (3165)
Event Date 10/27/2015
Event Type  Injury  
Event Description
During implantation, the surgeon was performing a percutaneous approach through a trocar and the head of the screw completely broke off.The shaft of the screw was unable to be removed and still remains in the patient's mandible.No secondary surgery is scheduled for screw removal.
 
Manufacturer Narrative
The screws were optically assessed and stereo microscopically investigated in the lab.The stereo microscopic investigation revealed a tensile crack due to mechanical overload.Further observation determined there were no indications of material or manufacturing defects discovered.Product device history records could not be reviewed because no lot number was provided.The results of the investigation conclude that the root cause for breakage was due to mechanical overload on the screws.If further information can be gathered that might add value to the contents of the investigation report, an additional follow-up report will be submitted.Correcting the mfr report number due to a typo this was originally submitted as 9610905-2055-00057 - 1 on 01/05/2016.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAXDRIVE MINI SCREW
Type of Device
BONE SCREW
Manufacturer (Section D)
KARL LEIBINGER MEDIZINTECHNIK GMBH & CO. KG
kolbinger strabe 10
muhlheim/donau, 78570
GM  78570
Manufacturer (Section G)
KARL LEIBINGER MEDIZINTECHNIK GMBH & CO. KG
kolbinger strabe 10
muhlheim/donau, 78570
GM   78570
Manufacturer Contact
jennifer damato
p.o. box 16369
jacksonville, FL 32245
9046417746
MDR Report Key5214764
MDR Text Key30979393
Report Number9610905-2015-00057
Device Sequence Number1
Product Code JEY
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K943347
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 10/28/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number25-872-11-91
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/27/2015
Initial Date FDA Received11/10/2015
Supplement Dates Manufacturer Received10/27/2015
Supplement Dates FDA Received05/05/2020
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age47 YR
-
-