Model Number SC-2316-50 |
Device Problems
Fracture (1260); High impedance (1291); Kinked (1339)
|
Patient Problem
Inadequate Pain Relief (2388)
|
Event Date 10/20/2015 |
Event Type
malfunction
|
Manufacturer Narrative
|
Additional suspect medical device component involved in the event: model #: sc-4316 lot #: 18182693 description: next generation anchor kit-sterile.
|
|
Event Description
|
A report was received that the patient underwent a lead explant procedure due to high impedances.During the procedure the physician noted that the lead appeared to be visually broken just below the clik anchor.The physician observed a deep kink on the lead, but did not observe any exposed wires.The patient was reportedly doing well post operatively.
|
|
Manufacturer Narrative
|
Sc-2316-50 (sn (b)(4)) the complaint has been confirmed.Visual and x-ray inspection of the lead revealed that multiple cables were completely broken at the bent/kinked location of the lead.The bent/kinked locations are 1 cm from both sides of the set screw mark of the clik anchor.There are no exposed cables at the fracture locations.Sc-4316 (ln 18182693) the clik anchor revealed no anomalies.
|
|
Event Description
|
A report was received that the patient underwent a lead explant procedure due to high impedances.During the procedure the physician noted that the lead appeared to be visually broken just below the clik anchor.The physician observed a deep kink on the lead, but did not observe any exposed wires.The patient was reportedly doing well post operatively.
|
|