Model Number SC-2316-50
Device Problems Fracture (1260); High impedance (1291); Kinked (1339)
Patient Problem Inadequate Pain Relief (2388)
Event Date 10/20/2015
Event Type  malfunction  
Manufacturer Narrative
Additional suspect medical device component involved in the event: model #: sc-4316 lot #: 18182693 description: next generation anchor kit-sterile.
 
Event Description
A report was received that the patient underwent a lead explant procedure due to high impedances.During the procedure the physician noted that the lead appeared to be visually broken just below the clik anchor.The physician observed a deep kink on the lead, but did not observe any exposed wires.The patient was reportedly doing well post operatively.
 
Manufacturer Narrative
Sc-2316-50 (sn (b)(4)) the complaint has been confirmed.Visual and x-ray inspection of the lead revealed that multiple cables were completely broken at the bent/kinked location of the lead.The bent/kinked locations are 1 cm from both sides of the set screw mark of the clik anchor.There are no exposed cables at the fracture locations.Sc-4316 (ln 18182693) the clik anchor revealed no anomalies.
 
Event Description
A report was received that the patient underwent a lead explant procedure due to high impedances.During the procedure the physician noted that the lead appeared to be visually broken just below the clik anchor.The physician observed a deep kink on the lead, but did not observe any exposed wires.The patient was reportedly doing well post operatively.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRECISION SPECTRA®
Type of Device
SPINAL CORD STIMULATOR
Manufacturer (Section D)
BOSTON SCIENTIFIC NEUROMODULATION
25155 rye canyon loop
valencia CA 91355
Manufacturer (Section G)
BOSTON SCIENTIFIC NEUROMODULATION
25155 rye canyon loop
valencia CA 91355
Manufacturer Contact
talar tahmasian
25155 rye canyon loop
valencia, CA 91355
6619494863
MDR Report Key5217143
MDR Text Key30985926
Report Number3006630150-2015-02957
Device Sequence Number1
Product Code LGW
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
030017
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 10/20/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date05/31/2017
Device Model NumberSC-2316-50
Device Catalogue NumberN/A
Device Lot NumberN/A
Other Device ID NumberM365SC2316500
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/02/2015
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/20/2015
Initial Date FDA Received11/11/2015
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received12/02/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/18/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-