• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: IRIS INTERNATIONAL IQ200; AUTOMATED URINE MICROSCOPY ANALYZER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

IRIS INTERNATIONAL IQ200; AUTOMATED URINE MICROSCOPY ANALYZER Back to Search Results
Catalog Number 700-3322
Device Problems Failure to Calibrate (2440); Device Operates Differently Than Expected (2913)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/24/2015
Event Type  malfunction  
Manufacturer Narrative
Iris field service engineer was sent to the customer location and found air in the fluidics system.The fse replaced the pipetter bypass valve p/n: 700-3880 and flushed the flow cell to resolve the issue.The fse reran calibration and controls, the system passed.The system was operational.(b)(4).
 
Event Description
The customer reported the calibration failed on their iq200 instrument.The customer also stated they were failing quality controls.No erroneous results generated or reported out of the lab.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IQ200
Type of Device
AUTOMATED URINE MICROSCOPY ANALYZER
Manufacturer (Section D)
IRIS INTERNATIONAL
9172 eton ave
chatsworth CA 91311
Manufacturer Contact
gopal mohanty
9172 eton avenue
chatsworth, CA 91311
8185277379
MDR Report Key5218681
MDR Text Key30992778
Report Number2023446-2015-00272
Device Sequence Number1
Product Code KQO
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K022774
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Health Professional
Type of Report Initial
Report Date 10/24/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number700-3322
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/24/2015
Initial Date FDA Received11/11/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/18/2008
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-