Carefusion file identification: (b)(4).Patient demographics such as age, date of birth, gender, and weight were not provided by the customer.The carefusion field service engineer (fse) went on site to adjust co2 analyzer but the system still would not calibrate.The fse came back and installed a new vmax encore 22 system at the customer¿s facility and performed system verification.The new unit conformed to the manufacturer¿s published performance specifications and was returned to normal operation.Carefusion continues to track and trend any incident related to this issue.(b)(4).
|
Failure analysis evaluation of the returned material: visual examination of the vmax module as received revealed no anomalies and physical damage.Testing of the unit continued by powering up the unit, and allowed to warm-up for 30 minutes.After 30 minutes, n2 calibration was performed, however the unit continued to fail.The c02 analyzer measured slope is < 0.5 or greater than 1.5.The slope on the lamp was too high.Investigation concludes that the co2 analyzer was determined to be faulty.The faulty c02 analyzer was replaced which resolved the reported issue.In conclusion, the likely cause of the event is attributed to the co2 analyzer.However, a definitive root cause is unknown.Carefusion continues to track and trend any incident related to this issue.
|